WiFi-500 Sensor Series User's Guide

Changing device logging settings

25

5. To change the sample rate, click

Sample Rate

, select a sample rate from the list, and click

Save

.

Figure 19: Sample Rate settings

6. To change the transmission period, click

Transmission Period

, select a transmission period from the list,

and click

Save

.

Figure 20: Transmission Period settings

7. To change the temperature/humidity alarms, click

Alarm Settings & Audit Check

, change any of the

following settings, and click

Save

.

o

Enable (

On

option) or disable (

Off

option) the low and high alarms for temperature and humidity (if

supported).

o

Enter alarm values for each enabled alarm. The low alarm value must be less than the high alarm
value.

o

To retain the alarm status when the temperature and/or humidity goes between the alarm range, click
the

On

option button for

Alarm Hold

.

For example, if the high alarm is set to 25 °C and the temperature rises to 26 °C, an alarm is triggered.
If the temperature then drops down to 21 °C and

Alarm hold

is selected, the device retains its high

alarm status.

If

Alarm hold

is not selected, the device reverts back to normal status, and the only way to check

missed alarms is to review the logged data.

o

To filter out unwanted alarms by delaying the alarm trigger, select a delay time from the

Alarm Delay

Alarm Trigger by

listbox.

To record an event on the graph when a device is physically checked, select the

Enable

Audit Check

checkbox. You indicate the device being checked by pressing and holding the device button for three
seconds.