Interactions
Phenolic substances (e.g. eugenol) may inhibit polymerization.
These materials must not be used in conjunction with Telio
Stains.

Usage
Telio Stains are used in conjunction with Telio CS C&B, Telio
CAD, Telio Lab and Telio Lab LC materials. The temporary
restorations are fabricated according to the corresponding
Instructions for Use.

Application in the dental practice

• Individualization of Telio CS C&B with Telio Stains

Before Telio Stains are applied, carefully blast the surfaces
(using Al

2

O

3

, 100 µm, at 1–2 bar/15–29 psi) or roughen with

coarse diamond burs. Thoroughly rinse with water and dry
with oil-free compressed air. Then wet the areas to be
bonded to extra-orally with Telio Activator. Brush Telio
Activator into the entire surface for at least 30 seconds using
an application brush in order to ensure an even distribution
and quick penetration of Telio Activator. Subsequently, allow
the Activator to react for another 30-60 seconds (total
reaction time 1–2 minutes). Now apply the Heliobond bon-
ding agent, thinly disperse it with blown air and polymerize
for 10 seconds (650 mW/cm

2

, e.g. bluephase using LOP).

Apply the Stains in very thin layers of max. 0.2 mm with a
brush or other suitable instrument (e.g. thin probe). The sha-
de intensity may be reduced and the consistency thinned by
mixing the stains with Heliobond. Finally, polymerize the
stains in accordance with the table below. Apply a suitable
varnish for temporaries according to the instructions of the
manufacturer.

• Polymerization times

Application in the dental laboratory

• Individualization of Telio CAD/Telio Lab with

Telio Stains
Reduce the restorations made of Telio CAD or Telio Lab in
the desired areas according to the principles of the cut-back

technique using cross-cut burs. Check the reduction (cut-
back) with the silicone key. Carefully blast the ground surfa-
ces ( Al

2

O

3

, 100 µm, 1–2 bar/15–29 psi). Then clean with

steam and dry with oil-free compressed air.
Distribute Telio Activator evenly and generously on the con-
ditioned surface and allow it to react for at least 2 minutes
to a maximum of 4 minutes.
Apply SR Composiv to the activated surface according to the
respective Instructions for Use and polymerize.
After that, the conditioned areas can be individualized with
Telio Stains. Apply Telio Stains in very thin layers of max.
0.2 mm with a brush or other suitable instrument or directly
from the thin metal cannula supplied. Final polymerization is
carried out in a light-polymerization device (see polymeriza-
tion table).

• Polymerization times for Telio Stains

The respective light-curing Telio components are pre-
polymerized to ensure that they stay in place and to be able
to apply a further layer. However, final polymerization of the
entire restoration is always required.

After polymerization, cover the Telio Stains with a layer of
Telio Lab or Telio Lab LC Transpa Incisal materials. For more
detailed information, please refer to the Telio Labside
Instructions for Use.

Important
Telio Stains should have room temperature during processing so
that a smooth consistency is ensured. Do not expose Telio

Polymeriza-

Manufacturer

Output

Polymeriza-

tion device

mW/cm

2

tion time

bluephase C8

Ivoclar Vivadent

800 mW/cm

2

15 s

bluephase

Ivoclar Vivadent

1200 mW/cm

2

10 s

bluephase 20i

Ivoclar Vivadent

2000 mW/cm

2

5 s

Polymeriza-

Manu-

Pre-

Final

tion device

facturer

polimerization

polimerization

Quick

Ivoclar

40 s

––

Vivadent

(SR Composiv
60 s)

Lumamat

®

100 Ivoclar

Exclusively use

11 min

(Targis Power

Vivadent

Quick for pre-

(program P2)

Upgrade)

polimerization

Spectramat

®

Ivoclar

5 min (Telio Stains 5 min

Vivadent

(2.5 min)

Visio Beta

1

3M ESPE

7 min, of which

15 min with

10 s vacuum

vacuum (pro-

(program depends gram depends
on type of device) on type of device

UniXS

1

Heraeus

90 s

2x 180 s

Solidilite EX

1

Shofu

1 min

5 min