Hoefer SP-2001 Vision Life Spectrophotometer User Manual

Page 34

background image



Version 1.0

Page 34






















Results Screen
Step 8
Insert the reference. Press 0A/100%T Key. This will be used for
all subsequent samples until changed.

Step 9
Insert sample and press

. This measures at the selected

wavelengths and displays the results. The ratio of wavelengths 1
and 2 absorbencies are calculated (both corrected by the
background wavelength value if selected). Gives concentration
based on absorbance at wavelength 1.
Repeat step 9 for all samples.

Press

to return to the Nucleic acid folder.


Press ::; to display available Options which are described
below.

Options (select using key pad numbers)
1. Return to parameters screen (step 1 above).
2. Print result via selected method.
3. Toggle graph on/off. The graph shows a wavescan plot

across the range 220 nm to 320 nm with cursors denoting
230, 260, 280 and (if background correction selected) 320
nm.

7. Sample number – add a prefix to the sample number and

reset the incrementing number to the desired value.

8. Save method – use the left and right arrows to select a folder

to store in (Favorites/Methods 1-9), press the down arrow
and enter name.

9. Auto-print – toggles auto-print on/off.

Exit options by pressing

, or wait.