Acquiring waveforms, Selecting ac-coupling – Fluke Biomedical 190M User Manual
Page 35

Using the Scope and Meter
Acquiring Waveforms
1
25
Acquiring Waveforms
Setting the Acquisition Speed and Waveform
Memory Depth
To set the acquisition speed, do the following:
1
SCOPE
Display the
SCOPE
key labels.
2
F4
Open the
WAVEFORM OPTIONS
menu.
3
ENTER
Select Acquisition:
Fast – for fast trace update rate;
shortest record length, decreased
zoom rate, no readings possible.
Full – maximum waveform detail;
10,000 samples per trace record
length, maximum zoom rate,
lower trace update rate.
Normal – optimal trace update
rate and zoom range combination
4
F4
Exit the menu
See also Table 2 in Chapter 8.
Selecting AC-Coupling
After a reset, the test tool is dc-coupled so that ac and dc
voltages appear on the screen.
Use ac-coupling when you wish to observe a small ac
signal that rides on a dc signal. To select ac-coupling, do
the following:
1
A
Display the
INPUT A
key labels.
2
F2
Highlight
AC
.
Observe that the bottom left of the screen displays the
ac-coupling icon:
.
You can define how Auto Set affects this setting, see
Chapter 6 “Changing the Auto Set Options.”