Automatic connect-and-view trigger, Edge trigger, Isolated external trigger (190m-2) – Fluke Biomedical 190M User Manual
Page 122

190M Series Medical ScopeMeter
Users Manual
112
Max. Delay ................................................... 60 s at 5 s/div
Automatic Connect-and-View Trigger
Source ............................................................... A, B, C, D
EXT
(190M-2)
Slope ............................................ Positive, Negative, Dual
Edge Trigger
Screen Update ........... Free Run, On Trigger, Single Shot
Source ....................................... A, B, C, D, EXT (190M-2)
Slope ........................................... Positive, Negative, Dual
Trigger Level Control Range ...........................
±
4 divisions
Trigger Sensitivity
DC to 5 MHz at >5 mV/div ........................... 0.5 divisions
DC to 5 MHz at 2 mV/div and 5 mV/div ........... 1 division
200 MHz (Fluke Biomedical190M-2) ................ 1 division
250
MHz
.........................................................
2
divisions
Isolated External Trigger (190M-2)
Bandwidth ................................................................. 3 kHz
Modes ...................................................... Automatic, Edge
Trigger Levels (DC to 3 kHz) ...................... 120 mV, 1.2 V