Automatic scope measurements, General, Dc voltage (vdc) – Fluke Biomedical 190M User Manual
Page 124: Ac voltage (vac)

190M Series Medical ScopeMeter
Users Manual
114
Automatic Scope Measurements
The accuracy of all readings is within
±
(% of reading +
number of counts) from 18
°
C to 28
°
C. Add 0.1x (specific
accuracy) for each
°
C below 18
°
C or above 28
°
C. For
voltage measurements with 10:1 probe, add probe
accuracy, see section ‘10:1 Probe’ on page 127. At least
1.5 waveform periods must be visible on the screen.
General
Inputs .............................................................. A, B, C and D
DC Common Mode Rejection Ratio (CMRR)......... >100 dB
AC Common Mode Rejection at 50, 60, or 400 Hz ....... >60 dB
DC Voltage (VDC)
Maximum Voltage
with 10:1 probe ...................................................... 1000 V
direct (1:1) .................................................................300 V
Maximum Resolution
with
10:1
probe
..........................................................
1 mV
direct (1:1) .............................................................. 100 µV
Full Scale Reading ............................................. 999 counts
Accuracy at 5 s to 10
μ
s/div
2
mV/div
..........................................
±
(1.5 % + 10 counts)
5 mV/div to 100 V/div .........................
±
(1.5 % + 5 counts)
Normal Mode AC Rejection at 50 or 60 Hz ............. >60 dB
AC Voltage (VAC)
Maximum Voltage
with
10:1
probe
....................................................
1000
V
direct
(1:1)
.............................................................
300
V
Maximum Resolution
with 10:1 probe ....................................................... 1 mV
direct (1:1) ........................................................... 100 µV
Full Scale Reading .......................................... 999 counts
Accuracy
DC
coupled:
DC
to
60
Hz
.................................
±
(1.5 % +10 counts)
AC coupled, low frequencies:
50
Hz
direct
(1:1)
........................
±
(1.5 % + 10 counts)
60
Hz
direct
(1:1)
........................
±
(1.9 % + 10 counts)
With the 10:1 probe the low frequency roll off point will
be lowered to 2 Hz, which improves the AC accuracy
for low frequencies. When possible use DC coupling
for maximum accuracy.
AC or DC coupled, high frequencies:
60 Hz to 20 kHz ..........................
±
(2.5 % + 15 counts)
20 kHz to 1 MHz ............................
±
(5 % + 20 counts)
1 MHz to 25 MHz .........................
±
(10 % + 20 counts)
For higher frequencies the test tool’s frequency roll off
starts affecting accuracy.
Normal Mode DC Rejection .............................. >50 dB