Temperature (temp), Dc voltage (vdc), Ac voltage (vac) – Fluke Biomedical 190M User Manual
Page 129: Ac+dc voltage (true rms), Amperes (amp)

Specifications
Meter Measurements for Model 190M-2
8
119
Temperature (TEMP)
With Optional Temperature Probe
Ranges (
°
C or
°
F) ................................... -40.0 to +100.0
°
-100.0
to
+250.0
°
-100.0
to
+500.0
°
-100
to
+1000
°
-100 to + 2500
°
Probe Sensitivity ............................. 1 mV/
°
C and 1 mV/
°
F
DC Voltage (VDC)
Ranges ...... 500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading......................................... 5000 counts
Accuracy .............................................
±
(0.5 % +5 counts)
Normal Mode AC Rejection at 50 or 60 Hz
±
1 % .... >60 dB
AC Voltage (VAC)
Ranges ...... 500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading......................................... 5000 counts
Accuracy
15 Hz to 60 Hz ..................................
±
(1 % +10 counts)
60 Hz to 1 kHz ...............................
±
(2.5 % +15 counts)
For higher frequencies the frequency roll off of the
Meter input starts affecting accuracy.
Normal Mode DC Rejection ................................... >50 dB
AC+DC Voltage (True RMS)
Ranges ..... 500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading ........................................ 5000 counts
Accuracy
DC
to
60
Hz
......................................
±
(1 % +10 counts)
60 Hz to 1 kHz ...............................
±
(2.5 % +15 counts)
For higher frequencies the frequency roll off of the Meter
input starts affecting accuracy.
All accuracies are valid if the waveform amplitude is
larger than 5 % of full scale.
Amperes (AMP)
With Optional Current Probe or Current Shunt
Ranges ............................. same as VDC, VAC, VAC+DC
Probe Sensitivity .................. 100 µV/A, 1 mV/A, 10 mV/A,
100 mV/A, 1 V/A, 10 V/A, and 100 V/A
Accuracy ........................... same as VDC, VAC, VAC+DC
(add current probe or current shunt accuracy)