Fluke Biomedical 190M User Manual
Page 16

190M Series Medical ScopeMeter
Users Manual
6
XW
Warning
•
Do not apply input voltages above the rating
of the instrument. Use caution when using
1:1 test leads because the probe tip voltage
will be directly transmitted to the test tool.
•
Do not use exposed metal BNC or banana
plug connectors. Fluke offers cables with
plastic, safety designed BNC connectors
suitable for the Medical ScopeMeter. See
Chapter 7, “Optional Accessories.”
•
Do not insert metal objects into connectors.
•
Use the test tool only as specified, or the
protection supplied by the test tool can be
compromised.
•
Carefully read all instructions.
•
Do not use the test tool if it operates
abnormally.
•
Do not use and disable the test tool if it is
damaged.
•
Keep fingers behind the finger guards on the
probes.
•
Use only correct Measurement Category
(CAT), voltage, and amperage rated probes,
test leads, and adapters for the measurement.
XW
Warning
•
Do not exceed the Measurement Category
(CAT) rating of the lowest rated individual
component of a test tool, probe, or
accessory.
•
Do not use the test tool around explosive
gas, vapor, or in damp or wet environments.
•
Measure a known voltage first to make sure
that the test tool operates correctly.
•
Examine the case before you use the test
tool. Look for cracks or missing plastic.
Carefully look at the insulation around the
terminals.
•
Do not work alone.
•
Comply with local and national safety codes.
Use personal protective equipment (approved
rubber gloves, face protection, and
flame resistant clothes) to prevent shock and
arc blast injury where hazardous live
conductors are exposed.
•
The battery door must be closed and locked
before you operate the test tool.