190M Series Medical ScopeMeter
Users Manual
120
Recorder
TrendPlot (Meter or Scope)
Chart recorder that plots a graph of min and max values of
Meter or Scope measurements over time.
Measurement Speed ......................... >5 measurements/s
Time/Div ............................................. 5 s/div to 30 min/div
Record Size (min, max, average) ...............
≥
18000 points
Recorded Time Span............................ 60 min to 22 days
...........................................................................................
Time Reference ...................... time from start, time of day
Scope Record
Records scope waveforms in deep memory while
displaying the waveform in Roll mode.
Source ...................................................... Input A, B, C, D
Max. Sample Speed (5 ms/div to 1 min/div) ...... 125 MS/s
Glitch capture (5 ms/div to 2 min/div) .......................... 8 ns
Time/Div in normal mode ................. 5 ms/div to 2 min/div
Record Size ....................................... 30k points per trace
Recorded Time Span................................. 6 s to 48 hours
Acquisition Modes ....................................... Single Sweep
Continuous
Roll
Start/Stop on Trigger
Time Reference ...................... time from start, time of day