UT4579 X-Band Up Converter

MN/UT4579.IOM

Maintenance and Troubleshooting

Rev. 0

76

Power

Supply

Reference

Oscillator

Fine

Step

Module

Sum

Loop

Module

Monitor & Control Assembly

RF Converter

Prime Power Cord Input

5/10 MHz Ref. Osc. Input

J2

P1

J1

J3

J4

J5

J7

J9

Summary Fault Relay Output

Serial Comm. Interface

(RS-485 / RS-232C)- COM 1

High Speed Bus (HSB)

IF Input (IF)

J6

J8

RF Loop Input (LP IN)

RF Loop Output (LP OUT)

RF Output (RF)

Transmit

I/O Switch

Module

(TSM)

IF Loop Input (LP IN)

IF Loop Output (LP OUT)

Figure 34. Converter Signal and Interconnecting Cable Diagram

(with TSM Switching Module).

D.2 M

AINTENANCE

T

ESTING

Use the instructions in Chapter 2, Installation, for installing the converter for checkout,
and the procedures in Chapter 3, System Operation, for operating the converter.
The converter is a Up Converter which translates the input IF frequency from 52 to 88
MHz (or optional 104 to 176 MHz) to an output RF frequency - for example, the RF
output frequency of the Model UT-4579 is 7900 to 8400 MHz. The IF input level is -35
dBm (typical), and the RF output level is +10 dBm at 1 dB compression.

D.2.1 T

EST

P

OINT

S

AMPLES

The IF input and RF output can be monitored at the RF Sample Test Points on the front
panel. A BNC connector is provided for the IF sample, and an SMA connector is
provided for the RF output. The RF sample output level is -20 dBc nominal, and the IF
sample input level is -20 dBc nominal.