3 solenoid valve principle, 8 maintenance – Bronkhorst EL-FLOW Base 2014 User Manual

Page 9

background image

BRONKHORST HIGH-TECH

1.7.3 Solenoid Valve Principle

The control valve used in the EL-FLOW Base series is a standard, direct operated control valve. It is a

normally closed solenoid valve. The plunger is lifted by the force of the magnetic field of the coil. The

diameter of the orifice under the plunger is optimised for the customer’s application.

flowcontrol

valve

The control valve is not designed to provide positive shut-off. It is recommended to install a separate

shut-off valve in the line if so required. Also pressure surges, as may occur during system pressurisation

must be avoided.

1.8

Maintenance

No routine maintenance is required to be performed on the controllers when they are used with clean

gas.

Units may be flushed with clean, dry inert gas.

In case of severe contamination it may be required to clean the inside of the instrument. After cleaning a

recalibration is needed. Bronkhorst High-Tech B.V. has a trained staff of servicemen available. Contact

your local supplier for cleaning and recalibration options.

If the equipment is not properly serviced, serious personal injury and/or damage to the equipment

could be the result. It is therefore important that servicing is performed by trained and qualified service

personnel.

9.17.061

EL-FLOW Base Series Mass Flow Controllers

Page 9