2 implementation class – Bronkhorst EL-FLOW Base 2014 User Manual

Page 18

background image

BRONKHORST HIGH-TECH

Step

Action

Indication

time

handling

1

Set instrument to

“bus config

mode”

both LEDs off

Press switch 5x briefly

2

Set tens of bus-

address

Green LED flashes

0.1 sec on,

0.1 sec off

count-flashes

start when switch

is pressed:

0.5 sec on,

0.5 sec off

time-out:

60 sec

Press switch and count green flashes

for tens of bus-address.

Release when wanted amount has

been count.

Counts up to max. 12 and than starts

at 0 again.

When counting fails, keep switch

pressed and restart counting for next

attempt.

3

Set units of bus-

Address

red LED flashes

0.1 sec on,

0.1 sec off

count-flashes

start when switch

is pressed:

0.5 sec on,

0.5 sec off

time-out:

60 sec

Press switch and count red flashes

for units of bus-address/MAC-ID.

Release when wanted amount has

been count.

Counts up to max. 9 and then starts

at 0 again.

When counting failed, keep switch

pressed and restart counting for next

attempt.

4

Set baud rate of

field bus

communication.

1 = 9600 Baud

2 = 19200 Baud

3 = 38400 Baud

both red and green

LED flashes

0.1 sec on,

0.1 sec off

count-flashes

start when switch

is pressed:

0.5 sec on,

0.5 sec off

time-out:

60 sec

Press switch and count red and green

flashes for baud rate setting.

Release when wanted amount has

been count.

Counts up to max. 3 and than starts

at 0 again.

When counting failed, keep switch

pressed and restart counting for next

attempt.

Note: selection of 0 means: No

change

Instrument returns to normal running/operation mode.

Changes are valid when they are made within the time-out times.

Actual setting can be checked by pressing the switch 3x briefly with intervals of max. 1 sec. for readout

the bus-address/MAC-ID and baud rate.

Note 1:

Value zero will be indicated by a period of 1 sec. off (0.5 sec. off + 0.5 sec. off).

When value zero is wanted, press switch shortly and release it again within 1 sec.

Note 2:

Before each action of flash-counting, the LED(s) to be used for counting will flash in a high frequency.

(Pattern: 0.1 sec on, 0.1 sec off). As soon as the switch is pressed-down, this LED (or both LEDs) will be

off and the counting sequence will start.

Note 3:

The parity setting cannot be read or changed using the micro-switch.

3.4.2 Implementation class

The physical and data link layer are implemented conforming to the "basic slave" implementation class

as described in document [2], “MODBUS over Serial Line specification and implementation guide V1.02”.

The following options have been implemented: