2 basic parameters – Bronkhorst EL-FLOW Base 2014 User Manual

Page 24

background image

BRONKHORST HIGH-TECH

3.7.2 Basic Parameters

Measured Value (Measure)

Type

Access

Range

FlowDDE

FlowBus

ModBus

Unsigned int

R

0…41942 8

1/0

0x0020/33

The measured value indicates the amount of mass flow metered by the instrument.

The signal of 0...100% will be presented in a range of 0...32000. The maximum signal to be expected is

131.07 %, which is: 41942.

Setpoint

Type

Access

Range

FlowDDE

FlowBus

ModBus

Unsigned int

RW

0…41942 9

1/1

0x0021/34

Setpoint is used to set the wanted amount of mass flow.

Signals are in the same range as the measured value, only setpoint is limited between 0 and 100 %.

Control Mode

Type

Access

Range

FlowDDE

FlowBus ModBus

Unsigned int

RW

0…18

12

1/4

0x0024/37

The Controller mode is used to select different functions of the instrument. The following modes are

available:

Value Mode

Instrument action

Setpoint source

0

DIGITAL_INPUT

Controlling

RS232/RS485

1

ANALOG_INPUT

Controlling

Analog input

3

VALVE_CLOSE

Valve closed

4

CONTROLLER_IDLE

Idle

5

TEST_MODE

Test mode enabled

7

SETPOINT_100

Controlling @100%

Fixed 100%

8

VALVE_OPEN

Valve full opened

9

CALIBRATION_MODE

Calibration mode enabled

12

SETPOINT_0

Controlling @0%

Fixed 0%

18

RS232_INPUT

Controlling

RS232 FlowBus

After power-up the control mode will always be set to DIGITAL_INPUT or ANALOG_INPUT, depending on

customer’s requirement. Check chapter 4.1.6, Changing Default Control Mode, to change the start-up

mode.

Page 24

EL-FLOW Base Series Mass Flow Controllers

9.17.061