1.6.4 Calibration

EL-FLOW Base instruments are Air calibrated. Bronkhorst High-Tech B.V. certifies that all instruments

meet the rated accuracy. They have been calibrated using measurement standards traceable to the

standards of the Dutch Metrology Institute (VSL).

The calibration is converted to the customer’s gas and conditions using a detailed conversion model.

This conversion adds a level of calibration uncertainty.

Thumb rule for calculating the conversion uncertainty is typical:

Uncertainty < 2% x CF

for CF > 1

Uncertainty < 2% / CF

for CF < 1

With CF defined as the approximate conversion factor, which can be calculated with:

2

1

2

1

ρ

ρ

=

p

p

C

C

CF

in which:

n

p

C

specific heat

n

ρ

density at normal conditions

(1)

calibration fluid (Air)

(2)

customer fluid

Check FLUIDAT® on http://www.fluidat.com

FLUIDAT® is a collection of routines to calculate physical properties of gases and liquids. These routines

are made available at the FLUIDAT® on the Net website.

EL-FLOW Base instruments are standard delivered without calibration certificate. Calibration certificates

can be ordered together with the instruments. Contact your local sales representative for more

information.

1.6.5 Features

Each instrument consists of an Analog interface, a digital RS-232 interface and a digital ModBus/RS485

interface. The analog and the digital interface can be used together at the same time.

According to the pin-designation both RS232 and Modbus/RS485 are assigned to the same pins. When

connecting these pins to either of the two, the instrument will automatically detect which protocol to

use.

Digital operation adds a lot of extra features (compared to analog operation) to the instruments.

Such as:

• setpoint slope (ramp function on setpoint for smooth control)

• direct reading at readout/control module or host computer

• several control/setpoint modes (e.g. purge/close valve)

• identification (serial number, model number, device type, user tag)

• adjustable controller settings for custom controller response