4 modbus rs485 operation – Bronkhorst EL-FLOW Base 2014 User Manual
Page 15

BRONKHORST HIGH-TECH
Examples of DDE client applications: FlowPlot, FlowView, MS-Office, LabView, Intouch, Wizcon.
The FLowDDE server also offers a lot of test facilities and user adjustable settings for efficient
communication with the connected flow controller.
How to setup a DDE link with FlowDDE is described in the help-file of the FlowDDE application.
Programming examples are available for making applications in: Visual Basic, LabView and Excel.
FLowDDE parameter numbers:
Reading/changing parameter values via FLowDDE offers the user a different interface to the instrument.
Besides the application name: ‘FLowDDE’ there is only need of:
• topic, used for channel number: ‘C(X)’
• item, used for parameter number: ‘P(Y)’
A DDE-parameter number is a unique number in a special FLowDDE instruments/parameter database
and not the same as the parameter number from the process on an instrument.
Node-address and process number will be translated by FLowDDE to a channel number.
FlowDDE and other Bronkhorst applications are available at the Bronkhorst download site:
A special RS232 cable (7.03.366) can be ordered separately. It consists of a T-part with 1 male and 1
female sub-D 9 connector on one instrument-side and a normal female sub-D 9 connector on the side of
the computer. By means of this cable it is possible to offer RS232 communication and still be able to
connect power-supply and analog interface through the (analog) sub-D 9 connector.
3.4
Modbus RS485 operation
This chapter is limited to the description of the interface between the Modbus Mass Flow Controller
with a master device. It will explain how to install an EL-FLOW Base instrument to your Modbus system.
It only contains the information that is needed most.
The implementation of the Modbus interface is based on the following standards:
[1] MODBUS Application Protocol Specification V1.1b, December 28, 2006
[2] MODBUS over Serial Line specification and implementation guide V1.02
There is no mutual communication between Modbus slaves; only between master and slave.
More detailed information about Modbus can be found at
or any website of
the (local) Modbus organisation of your country (when available).
Base
FlowDDE
server
Application 1
Application 2
Application n
RS232
DDE-link
DDE-link
DDE-link
Windows based Personal Computer
Instrument
9.17.061
EL-FLOW Base Series Mass Flow Controllers