3 fluid information, 4 auto zeroing – Bronkhorst EL-FLOW Base 2014 User Manual

Page 26

background image

BRONKHORST HIGH-TECH

Customer model

Type

Access

Range

FlowDDE

FlowBus

ModBus PDU

Unsigned char[16]

RW 

-

93

113/4

0xF120..0xF127/61729..61736

Digital instrument manufacturing configuration information string.

This string can be used by Bronkhorst High-Tech to add extra information to the model number

information.

4.1.3 Fluid Information

Next parameters give information about the fluid range of the instrument.

Fluid name

Type

Access

Range

FlowDDE

FlowBus

ModBus PDU

Unsigned char[10]

RW 

-

25

1/17

0x8188..0x818C/33161..33165

Fluid name consists of the name of the fluid. Up to 10 characters are available for storage of this name.

Fluid unit

Type

Access

Range

FlowDDE

FlowBus

ModBus PDU

Unsigned char[7]

R

-

129

1/31

0x81F8..0x81FB/33273..33276

The Fluid unit can be read by parameter ‘capacity unit’. This parameter contains the unit in maximal 7

characters.

Fluid Capacity (@100%)

Type

Access

Range

FlowDDE

FlowBus

ModBus PDU

Float

R

±1E-10… ±1E+10

21

1/13

0x8168..0x8169/33129..33130

Capacity is the maximum value (span) at 100% for direct reading in sensor base units.

Fluid Capacity (@0%)

Type

Access

Range

FlowDDE

FlowBus

ModBus PDU

Float

R

±1E-10… ±1E+10

183

33/22

0xA1B0..0xA1B1/41393..41394

This is the capacity zero point (offset) for direct reading in sensor base units.

4.1.4 Auto Zeroing

To start the auto zero-procedure two parameters should be written:

Control Mode

Type

Access

Range

FlowDDE

FlowBus ModBus

Unsigned int

RW

0…18

12

1/4

0x0024/37

Check chapter 3.7.2, “Basic Parameters”, for available control modes.

Calibration Mode

Type

Access

Range

FlowDDE

FlowBus ModBus

Unsigned int

RW 

9

58

115/1

0x0E61/3682

Value Mode

Instrument action

0

IDLE

Idle

9

AUTO_ZERO

Auto-zeroing

255

ERROR

Idle

Procedure:

Step 1: Set Control Mode to CALIBRATION_MODE (9)

Step 2: Set Calibration Mode to AUTO_ZERO(9)

Step 3: Check Calibration Mode,

IDLE

Auto-zeroing succeeded

AUTO_ZERO Auto-zeroing active

ERROR

Auto-zeroing failed

Page 26

EL-FLOW Base Series Mass Flow Controllers