BRONKHORST HIGH-TECH

Diagnostics (08)
The following sub-functions are supported:

Sub-function code (dec)

Name

00

Return Query Data

10

Clear Counters and Diagnostics Register

11

Return Bus Message Count

12

Return Bus Communication Error Count

13

Return Bus Exception Error Count

14

Return Slave Message Count

15

Return Slave No Response Count

16

Return Slave NAK Count (always 0)

17

Return Slave Busy Count (always 0)

18

Return Bus Character Overrun Count

!

The maximum message size for the Return Query Data sub function is 100 bytes at 9600 baud (200

bytes at 19200 baud and 400 bytes at 38400 baud). When this size is exceeded, corrupted responses

may be received.

Possible exception responses:

• 01, ILLEGAL FUNCTION, in case of not-supported sub-function

• 03, ILLEGAL DATA VALUE, in case of an incorrect value for the data field
Report Slave ID (17)
The Slave ID field in the response is a string with the same contents as FlowDDE parameter 1 (indent

number + version nr/serial nr). The Run Indicator Status field in this message will indicate ON when the

device is in normal operating mode (FB_NORMAL).

Possible exception responses:

• 04, SLAVE DEVICE FAILURE, in case of an internal error

3.4.5 Available parameters

Modbus registers (in the data model) are numbered from 1 to 65536. In a Modbus PDU (Protocol Data

Unit) these registers are addressed from 0 to 65535.

The following table lists the most commonly used parameters.

MODBUS REGISTERS

PARAMETER NAME

PARAMETER

TYPE

ACCESS

PDU ADDRESS

hex

REGISTER NUMBER

REMARK

Hex

Dec

Wink

Unsigned char

W

0x0000

0x0001

1

Value 14592

Init/reset

Unsigned char

RW

0x000A

0x000B

11

Valve output

Unsigned int

RW

0x001F

0x0020

32

0..32767

Measure

Unsigned int

R

0x0020

0x0021

33

Setpoint

Unsigned int

RW

0x0021

0x0022

34

Setpoint slope

Unsigned int

RW

0x0022

0x0023

35

Analog input

Unsigned int

R

0x0023

0x0024

36

Setp. control modes

Unsigned char

RW

0x0024

0x0025

37

Sensor type

Unsigned char

RW 

0x002E

0x002F

47

Capunit

Unsigned char

RW 

0x002F

0x0030

48

Fluid number

Unsigned char

RW

0x0030

0x0031

49

Alarminfo

Unsigned char

R

0x0034

0x0035

53

Temperature

Unsigned int

R

0x0427

0x0428

1064

Identnumber

Unsigned char

RW 

0x0E2C

0x0E2D

3629

ContrResp

Unsigned char

RW 

0x0E45

0x0E46

3654

CycleTime

Unsigned char

R

0x0E4C

0x0E4D

3661

Page 20

EL-FLOW Base Series Mass Flow Controllers

9.17.061