BRONKHORST HIGH-TECH

Symptom

Possible cause

Action

Small flow at zero setpoint

Valve leaks due to damaged plunger or dirt in

orifice

6a) clean orifice and/or, when replacing

plunger assembly, see ‘1e’

Pressure too high or much too low

6b) apply correct pressure

High flow at zero setpoint

Damaged diaphragm (only applicable to

valves with membrane)

7a) replace membrane seal

Page 32

EL-FLOW Base Series Mass Flow Controllers

9.17.061