4 Advanced Operation

4.1

Reading and Changing Instrument Parameters

4.1.1 Introduction

!

All parameters described in this chapter have influence on the behaviour of the mass-flow meter. Please

be aware that wrong settings can disorder the output and control response.

To avoid un careless changes of these parameters, these parameters are locked. To un-lock these

parameters use set parameter “Init Reset” to “UN-LOCKED”

Init Reset

Type

Access

Range

FlowDDE

FlowBus

ModBus

Unsigned char

RW

82/64

7

0/10

0x000A/11

The Init Reset parameter is used to ‘Un-Lock’ advanced parameters for writing. This parameter knows

the following values:

Value Mode

Instrument action

82

LOCKED

Advanced parameters are

read-only

64

UN_LOCKED

Advanced parameters are

write- en readable.

This parameter is always set to “LOCKED” at power-up.

4.1.2 Identification

Serial number

Type

Access

Range

FlowDDE

FlowBus

ModBus

Unsigned char[20]

R

-

92

113/3

0xF118..0xF11F/61721..61728

This parameter consists of a maximum 20-byte string with instrument serial number for identification.

Example: “M0202123A”

BHT Model number

Type

Access

Range

FlowDDE

FlowBus

ModBus PDU

Unsigned char[14]

R

-

91

113/2

0xF111..0xF117/61713..61719

Bronkhorst High-Tech instrument model number information string.

Firmware version

Type

Access

Range

FlowDDE

FlowBus

ModBus PDU

Unsigned char[5]

R

-

105

113/5

0xF128..0XF12A/61737..61739

Revision number of firmware. Eg. “V1.12”

Usertag

Type

Access

Range

FlowDDE

FlowBus

ModBus PDU

Unsigned char[13]

RW 

-

115

113/6

0xF130..0xF136/61745..61751

User definable alias string. Maximum 13 characters allow the user to give the instrument his own tag

name.

It is advised here to limit the name up to 7 characters when using E-7000 readout and control modules.

These modules can display the tag name of an instrument only up to 7 characters.

Eg.: “Room1s6”

9.17.061

EL-FLOW Base Series Mass Flow Controllers