7 basic parameters and properties, 1 introduction – Bronkhorst EL-FLOW Base 2014 User Manual

Page 23

background image

BRONKHORST HIGH-TECH

Green LED

Time

Indication

0.2 sec off

Instrument is busy performing any special function.

E.g. auto-zero or self-test

Green LED indication modes (no switch used)

Red LED

Time

Indication

Off

Continuously

No error

Flash

Variable

Bus activity on the Modbus interface

On

Continuously

Critical error message

A serious error occurred in the instrument

Instrument needs service before further using

Red LED indication modes (no switch used)

3.7

Basic Parameters and Properties

3.7.1 Introduction

Every parameter has its own properties. These properties are given in a table as shown:

Type

Access

Range

FlowDDE

FlowBus

ModBus

[type]

RW 

[x]…[y]

[FB]

[Pro]/[Par] [address]/[index]

Type

Unsigned char

1 byte character

Unsigned char[x] x byte array (string)

Unsigned int

2 byte unsigned integer

Float

4 byte floating point

Access

R

The parameter is read-only

RW

The parameter can be read and write

RW The parameter can only be written when the Init Reset parameter is set to 64. See Chapter

4.1.1, General Product Information for more details.

Range

Some parameters only accept values within a certain range:

[x]

Minimal value of the range.

[y]

Maximal value of the range.

FlowDDE

Parameter number in FlowDDE. Check chapter 3.3, “Basic RS232 Flowbus operation”, for detailed

information.

FlowBus

Process and parameter number to address parameters using the FlowBus protocol.

[Pro]

Flowbus process number

[Par]

Flowbus parameter number

Check document 9.17.027 , “RS232 interface with FLOW-BUS protocol”, for detailed information.

ModBus

PDU Address and register number to address parameters using the ModBus protocol.

[address] Hexadecimal PDU address .

[index]

Decimal register number.

For the ModBus protocol every 2 bytes are addressed separately. Check chapter 3.4, “Modbus RS485

operation” for more details.

9.17.061

EL-FLOW Base Series Mass Flow Controllers

Page 23