beautypg.com

3 calibration, Table 5–2. potentiometer adjustment procedures -3, Figure 5–1. potentiometer locations -3 – AMETEK DLM Series User Manual

Page 73

background image

DLM-E 3kW & 4kW Series Power Supplies

Maintenance and Troubleshooting

5.3 Calibration

Calibration of the output voltage, current, or OVP is accomplished using multiturn trimpots.
Table 5–2 gives the location, function, and effect of each potentiometer.

Calibration is performed at the factory during testing. Recalibration should be performed
annually and following major repairs. With the cover in place, calibration should be done
through access holes in the cover or front panel. See Figure 5–1.

Figure 5–1. Potentiometer Locations

Location

Function

Adjustment Procedure

Top Cover

Output Voltage

Clockwise rotation increases output voltage setting

Top Cover

Output Current

Clockwise rotation increases output current setting

Top Cover

Front Panel Voltmeter

Clockwise rotation increases voltage meter reading

Top Cover

Front Panel Current Meter Clockwise rotation increases current meter reading

Front Panel OVP

Clockwise rotation increases OVP shutdown setting

Table 5–2. Potentiometer Adjustment Procedures

M362000-01 Rev E

5-3