DLM-E 3kW & 4kW Series Power Supplies

Installation

M362000-01 Rev E

2-7

16. OTEMP (Over Temperature) indicator: Red LED lights when an over temperature

shutdown has occurred.

17. LOCK (Lockout) indicator: Green LED lights when in Lockout mode. Activated by

connection through rear panel DIP switch. Can only be activated when the front panel
LOCAL/REMOTE switch is in the REMOTE position. Once the Lock function has been
activated, it disables LOCAL control for all output control functions except the AC power
switch, which remains under front panel control.

The following indicators will be illuminated only when the optional GPIB controller is installed:

18. ERR (Error) indicator: Red LED lights to signal a GPIB programming error has occurred.

19. SRQ (Service Request) indicator: Green LED lights to signal GPIB service request by

the supply.

20. ADDR (Address) indicator: Green LED lights when the unit is addressed by a remote

controller.

2.4.2

Rear Panel

Refer to Figure 2–2 or Figure 2-3 (depending on the model) and the descriptions below.

1. Positive Output (+). Bus bar for 5V through 80V models.

TB2–1 and 2 for 150V through 600V models.

2. Negative Output (). Bus bar for 5V through 80V models.

TB2–3 and 4 for 150V through 600V models.

3. Output connector location for 150V through 600V models.

4. Programming and Monitor Connector, J3. I/O connector for input programming and

analog output monitoring signals as well as status indication and remote shutdown
signals. See Table 4–1 for individual pin descriptions.

5. DIP

Switch

S1. Eight–position right angle slide DIP switch. Controls full scale settings for

Voltage, Current and OVP programming range, Voltage and Current Output Monitor
range, Remote On–Off logic selection, Master/Slave operation and Lockout operating
mode selections. See Table 4–2 for Rear Panel DIP switch functions and settings.

6. Parallel Port connectors, J12 and J13. Used in conjunction with S1 setting to control

multiple units in parallel mode with current sharing. See Table 4–2 for Rear Panel DIP
switch functions and settings and Table 4–3 for parallel port connector function and
pinout.

7. Sense

connector,

J11. Remote sense lead connection for local and remote load voltage

sensing on all models. Connections shown are for local sensing on all models.

8. AC Input chassis safety ground stud.

9. AC Input Connector/fuse block with removable safety cover.

10. Area for optional GPIB assembly.