Advanced Operation

DLM-E 3kW & 4kW Series Power Supplies

connector configurations and connections required for remote programming of output voltage
and/or current limit without using the REMOTE/LOCAL switch.

Remote Programming Options

Control of ...

Programming Scales *

Programming with the
REMOTE/LOCAL Switch

Output Voltage
Current Limit
Over Voltage Setting

0–5V, 0–10V, 0–5K

Programming without the
REMOTE/LOCAL Switch
(Jumper J3 Connector)

Output Voltage and/or
Current Limit

0–5V, 0–10V, 0–5K
Local control

* These scales may be used in any combination

4.3.1

Programming Output Voltage and Current Limit with the
REMOTE/LOCAL Switch

The front panel REMOTE/LOCAL Programming switch will allow you to switch back and forth
between remote and local operation when programming output voltage and current limit with
external voltage and/or current sources.

For programming output voltage and current limit using the REMOTE/LOCAL switch:

1. Connect a programming source between pins 9 (voltage programming input/positive)

and 12 (return).

2. Connect a programming source between pins 10 (current limit programming

input/positive) and 12 (return).

3. Connect a programming source between pins 3 (OVP programming input/positive) and

12 (return).

4. Connect a TTL high signal to Pin 1 (remote output enable). An external source must be

referenced to Pin 6 (common). Pin 15 (+5V) may be used as a source.

5. Set the front panel REMOTE/LOCAL switch to REM.