Specifications, Built-in touch screen computer, Cameras and lenses – UVP 81-0345-01 ChemiDoc-ItTS2 Imager User Manual
Page 5: Ethidium bromide (etbr) emission filter

GelDoc-It
TS2
and ChemiDoc-It
TS2
Imagers
5
Specifications
Power Requirements:
100/115V, 50/60Hz; 3.1 Amps at 120 Volts
230V, 50/60Hz; 1.55 Amps at 230 Volts
Mains supply voltage fluctuations are not to exceed 10 percent of the nominal
supply voltage
Pollution Degree:
2
Installation Category:
II
Altitude:
Up to 2000m
Ambient Temperature:
5°C to 40°C
Humidity:
Maximum relative humidity of 80% for temperatures up to 31°C, decreasing
linearly to 50% maximum relative humidity at 40°C
Built-In Touch Screen Computer
Operating System:
Windows
®
7, 64-Bit
Connectivity Ports:
2 USB (side of system)
4 USB, VGA and Ethernet (rear of system)
Wireless Networking Capability:
802.11 b/g/n
USB Flash Drive Capacity:
2 GB (minimum)
Internal Hard Drive:
250 GB (minimum)
Software:
TS2 (Touch Screen) Software
Cameras and Lenses
GelDoc-It
TS2
: The GelDoc-It
TS2
is equipped with the GelCam 310 camera, a scientific-grade monochrome
CCD camera with a resolution of 2.0MP (1600x1200). The GelCam 310 is equipped with a 12.5-75mm f/1.2
optical zoom lens.
ChemiDoc-It
TS2
: The ChemiDoc-It
TS2
is equipped with either the MegaCam 810 or BioChemi™ 510
scientific-grade monochrome CCD camera. Both cameras are Peltier cooled and offer full 16-bit file bit
depth:
The BioChemi 510 camera is a 2.1MP unit with Peltier cooling to 35°C below ambient. The
BioChemi 510 has a peak quantum efficiency of 50% and is capable of binning from 1x1 to 8x8.
This camera is equipped with a 12.5-75mm f/1.2 optical zoom lens.
The MegaCam 810 camera is an 8.1MP camera (3296x2472) with Peltier cooling to 35°C below
ambient. The MegaCam 810 has a peak quantum efficiency of 50% and is also capable of binning
from 1x1 to 8x8. This camera is equipped with either a 50mm f/1.2 lens or a 30mm f/1.4 lens, both
of which have a fixed focal length (non-optical zoom capable).
All camera settings are factory pre-set for optimum performance when viewing gels, films or membranes
under low light level conditions. Contact UVP Technical Support prior to making any adjustments to internal
camera settings.
Ethidium Bromide (EtBr) Emission Filter
The ethidium bromide (50mm
2
) UV-blocking bandpass interference filter blocks UV and IR radiation emitted
from the transilluminator. The filter is placed in the slide-out filter wheel assembly on the side of the Imager.
The filter allows visualization of fluorophores from 580-630nm, targeting the ethidium bromide emission peak
of 605nm.
Additional filters are available for other specific fluorophores, including custom filters. Filters can also be
removed when imaging non-fluorescent media (including chemiluminescent blots, protein gels, colony
plates, etc.) in order to produce brighter images. Contact UVP for ordering information.