Inspect the thin wire within each glass fuse to see if there is a break in the wire. If so, replace the
fuse(s). If fuses are blowing repeatedly, contact UVP Technical Support for additional
troubleshooting.

Transilluminator Will Not Turn On

1. In addition to turning on “UV transillumination” in the Lighting menu within the TS2 Software, the

transilluminator itself has a power switch. Make sure that the green transilluminator power switch,
located on the front of the transilluminator, is in the ON position.

2. For UV protection and to extend the life of the UV transilluminator, the system incorporates a

customizable transilluminator shutoff timer built into the software. For additional information, refer to
the Lighting section of this manual.

Error Messages Appear on the Screen

1. An error message that is related to the TS2 software interface or Microsoft Windows may appear

on the screen. If the message is related to Microsoft Windows, such as a reminder to activate or
update the copy of Windows, please contact your system administrator for assistance.

2. If an error message appears repeatedly and your system administrator does not recognize it as a

Microsoft Windows error, contact UVP Technical Support for further assistance.


Care and Cleaning

Use only mild soap or detergent solution for cleaning the GelDoc-It

TS2

/ChemiDoc-It

TS2

. Do NOT use oil- or

petroleum-based cleaners for the cabinet. Ensure that the system is turned OFF and unplugged during
cleaning.

When cleaning the transilluminator surface, use a damp soft cloth or sponge. Never use abrasive cleaners
which can damage the UV filter surface.


Technical Support

UVP offers free lifetime technical support on all of its products and software. Should you have any questions
regarding the product’s use, operation or repair, contact UVP’s offices at the locations below, or visit
www.uvp.com.

If you are in North America, South

America, East Asia or Australia:

If you are in Europe, Africa, the

Middle East of Western Asia:

Call (800) 452-6788 or (909) 946-

3197, and ask for Customer Service

during regular business days, between

7:00 am and 5:00 pm, PST.

Call +44(0) 1223-420022, and ask for

Customer Service during regular

business days between 9:00 am and

5:30 pm.

E-mail your message to:

[email protected]

E-mail your message to: [email protected]

Fax Customer Service, and send it to

(909) 946-3597

Fax Customer Service, and send it to:

+44(0) 1223-420561

Write to: UVP, LLC 2066 W. 11

th

Street, Upland, CA 91786 USA

Write to: Ultra-Violet Products Ltd

Unit 1, Trinity Hall Farm Estate, Nuffield

Road, Cambridge CB4 1TG UK




GelDoc-It, ChemiDoc-It, VisionWorks and FirstLight are registered trademarks of UVP, LLC.
BioChemi, BioLite and Visi-Blue are trademarks of UVP, LLC.

All other trademarks are the property of their respective owners.

This manual is related to the following products: