beautypg.com

Troubleshooting – SOR Single Point Ultrasonic Level Switch User Manual

Page 14

background image

14/16

Form 829 (08.13) ©SOR Inc.

Model No.

of 1 pt control

for metal notch

integral sensor

Replacement

Board

includes

hardware

Model No.

of 1 pt control

for metal notch

remote sensor

Replacement

Board

includes

hardware

Model No.

of 1 pt control

for viscous

sensor

Replacement

Board

includes

hardware

711K1-N-P6-C

5002-996

711R1-N-P6-C

5002-958

711K1-V-P6-C

5003-021

711K1-N-P7-C

5002-997

711R1-N-P7-C

5002-959

711K1-V-P7-C

5003-022

711K1-N-P8-C

5002-998

711R1-N-P8-C

5002-960

711K1-V-P8-C

5003-023

721K2-N-P6-C

5002-999

721R2-N-P6-C

5002-961

721K2-V-P6-C

5003-024

721K2-N-P7-C

5003-001

721R2-N-P7-C

5002-962

721K2-V-P7-C

5003-025

721K2-N-P8-C

5003-002

721R2-N-P8-C

5002-963

721K2-V-P8-C

5003-026

Model No.

of 2 pt control

for metal notch

integral sensor

Replacement

Board

includes

hardware

Model No.

of 2 pt control

for metal notch

remote sensor

Replacement

Board

includes

hardware

Model No.

of 1 pt remote

control for

viscous sensor

Replacement

Board

includes

hardware

712K3-N-P6-C

5003-003

712R3-N-P6-C

5002-964

711R1-V-P6-C

5003-015

712K3-N-P7-C

5003-004

712R3-N-P7-C

5002-965

711R1-V-P7-C

5003-016

712K3-N-P8-C

5003-005

712R3-N-P8-C

5002-966

711R1-V-P8-C

5003-017

722K3-N-P6-C

5003-006

722R3-N-P6-C

5002-967

721R2-V-P6-C

5003-018

722K3-N-P7-C

5003-007

722R3-N-P7-C

5002-968

721R2-V-P7-C

5003-019

722K3-N-P8-C

5003-008

722R3-N-P8-C

5002-969

721R2-V-P8-C

5003-020

Symptom

Possible Causes

Solutions

Sensor
Monitor
relay
de-energized
(Model 721,
722 only)

No power to unit

Restore power.

One or both sensor coax cable(s) unplugged
from PC board. Check sensor housing on
remote units.

Replug coax cables into the board according
to the Field Replacement procedure (page 11).

Internal or external damage to sensor.
Sensor damage can be verified using sensor
test procedure (page 13). If sensor tests
okay, replace circuit board according to field
replacement procedure (page 11).

Replace entire instrument or contact factory
for replacement sensor. The factory will
request model and serial numbers to provide
replacement sensor.

No response
to liquid level

No power to unit

Restore power.

Blown fuse

Replace fuse (page 13)

One or both sensor cable(s) unplugged
from PC board. Check sensor housing on
remote units.

Replug sensor cables into the board
according to the Field Replacement procedure
(page 11).

Internal or external damage to sensor.
Sensor damage can be verified using sensor
test procedure (page 8). If sensor tests
okay, replace circuit board according to field
replacement procedure (page 11).

Replace entire instrument or contact factory
for replacement sensor. The factory will
request model and serial numbers to provide
replacement sensor.

Troubleshooting

NOTE: Agency certifi ed units, (FM, CSA, IEC) must be returned to SOR for repairs.