Preface

P a r r I n s t r u m e n t C o m p a n y

6

Frequency: Calorimeters can be operated from
either a 50 or 60 Hertz power supply without affect-
ing their operation or calibration.
Current: The total current drawn should not exceed
the rating shown on the data plate on the calorim-
eter by more than 10 percent.

Environmental Conditions

Operating: 15 ºC to 30 ºC; maximum relative humid-
ity of 80% non-condensing. Installation Category II
(over voltage) in accordance with IEC 664. Pollution
degree 2 in accordance with IEC 664.
Altitude Limit: 2,000 meters.

Storage: -25 ºC and 65 ºC; 10% to 85% relative
humidity.

Provisions for Lifting and Carrying

Before moving the instrument, disconnect all con-
nections from the rear of the apparatus. Lift the
instrument by grabbing underneath each corner.

Cleaning & Maintenance

Periodic cleaning may be performed on the exterior
surfaces of the instrument with a lightly dampened
cloth containing mild soap solution. All power
should be disconnected when cleaning the instru-
ment. There are no user serviceable parts inside the
product other than what is specifically called out
and discussed in this manual. Advanced trouble-
shooting instructions beyond the scope of this
manual can be obtained by calling Parr Instrument
Company in order to determine which part(s) may
be replaced or serviced.

Getting Started

These steps are offered to help the user become
familiar with, install, operate and develop the full
capabilities of the Parr 6200 Calorimeter.

1. Review the Concept of Operations, Chapter 1, to

get an understanding of the overall capabilities
of the calorimeter and microprocessor control.

2. Unpack and install the calorimeter in accordance

with Installation, Chapter 2. This simple, step-
wise procedure will acquaint the user with the

various parts of the calorimeter and make it
easier to understand the operating instructions
which follow.

3. Turn the power switch ON (located on the back).

Turn to the Instrument Description, Chapter 3, to
review the touch screen controls.

4. Review the Program Installation and Control,

Chapter 4, to match the factory settings to the in-
tended mode of operation. Any required changes
can be made to the program parameters located
in the Main Menu.

5. Review the Reporting Instructions, Chapter 7, to

become familiar with the manner in which calo-
rimetry corrections are entered. Also discussed
are generating final reports, editing and clearing
memory.

6. Turn to the Menu Operating Instructions, Ap-

pendix A, to review the menu functions used
to modify the program contained in the 6200
Calorimeter. A review of the menus will provide
a good idea of the capabilities and flexibility
designed into this instrument.

7. Review the Calculations, Appendix B. This pro-

vides information about calculations performed
by the 6200 Calorimeter.

8. Review Standardization, Appendix C. This will

serve two important functions. First, it provides
instructions on generating the energy equivalent
factor required to calculate the heat of combus-
tion of unknown samples. Secondly, it will give
the user the opportunity to run tests on a mate-
rial with a known heat of combustion to become
familiar with the instrument and confirm that the
instrument and operating procedures are pro-
ducing results with acceptable precision. Most
6200 Calorimeters will have an energy equiva-
lent of approximately 2400 calories per ºC. The
runs for standardization and determinations are
identical, except for the setting of the instrument
to the standardization or determination mode.

9. Review the Communication Interfaces, Appendix

D, for the correct installation of any peripherals
connected to the 6200 Calorimeter.

10. After successful standardization, the 6200 Calo-

rimeter should be ready for testing samples.