Table c-3, Calorimeter control limit values in btu/lb, Standardization – Parr Instrument 6200 User Manual

Page 60

background image

C

Standardization

P a r r I n s t r u m e n t C o m p a n y

58

Table C-3

Calorimeter Control Limit Values in BTU/lb When Benzoic Acid is Used as a Test Sample

Accepted heat of combustion taken as 11373 BTU/lb.
Instrument precision 0.10% RSD.
Control limits based on 99% confidence (3 sigma) values.
Values are in BTU/lb.

NUMBER OF

OBSERVATIONS

IN A GROUP

UCL FOR THE RANGE

(HIGH – LOW) WITHIN

THE GROUP

UCL FOR THE

RSD WITHIN

THE GROUP

MAXIMUM PERMISSIBLE

DEVIATION OF THE GROUP MEAN

FROM THE ACCEPTED VALUE OR

GRAND MEAN

1

34.1

2

41.9

0.261%

24.1

3

49.6

0.228%

19.7

4

53.4

0.209%

17.1

5

55.9

0.196%

15.3

6

57.8

0.187%

13.9

7

59.2

0.181%

12.9

8

60.4

0.175%

12.1

9

61.3

0.171%

11.4

10

62.2

0.167%

10.8

11

62.9

0.164%

10.3

12

63.6

0.161%

9.8

13

64.2

0.159%

9.5

14

64.8

0.156%

9.1

15

65.3

0.154%

8.8

16

65.8

0.153%

8.5

17

66.2

0.151%

8.3

18

66.6

0.150%

8.0

19

67.0

0.148%

7.8

20

67.4

0.147%

7.6

21

67.7

0.146%

7.4

22

68.0

0.145%

7.3

23

68.3

0.144%

7.1

24

68.6

0.143%

7.0

25

68.9

0.142%

6.8