Program installation & control – Parr Instrument 6200 User Manual
Page 21

Program Installation & Control
6200
4
w w w . p a r r i n s t . c o m
19
Calculation Factors
Nitric Acid Factor
1.58
Acid Multiplier
0.0709
Sulfur Value is Percent
ON
Sulfur Multiplier
0.6238
Fuse Multiplier
1.0
Use Offset Correction (ISO)
OFF
Offset Value
0.0
Data Entry Controls
Prompt for Bomb ID
ON
Weight Entry Modes
Touch Screen
Acid Entry Mode
Touch Screen
Net Heat Entry Modes
Touch Screen
Auto Sample ID Controls
ON
Sample Weight Warning above
2.0
Spike Weight Entry Mode
Touch Screen
Sulfur Entry Mode
Touch Screen
Moisture Entry Mode
Touch Screen
Auto Preweigh Controls
ON
Auto Sample ID Controls
Automatic Sample ID
ON
Automatic Sample ID Increment
1
Automatic Sample ID Number
1
Auto Preweigh Controls
Automatic Preweigh ID
ON
Automatic Preweigh ID Increment 1
Automatic Preweigh ID Number
1
Reporting Controls
Report Width
40
Automatic Reporting
ON
Auto Report Destination
Printer
Individual Printed Reports
OFF
Edit Final Reports
OFF
Recalculate Final Reports
OFF
Use New EE Values in Recalculation OFF
Communication Controls
Printer Type
Parr 1758
Balance Port
Network Interface
Printer Destination
Local (USB)
Bar Code Port
Network Data Devices
Balance Port Communications
Balance Type
Generic
Balance Port Device
Customize Balance Settings
Balance Port Settings
Number of Data Bits
8
Parity
None
Number of Stop Bits
1
Handshaking
None
Baud Rate
9600
Data Characters from Balance
8
Data Precision
4
Transfer Timeout (seconds)
10
Balance Handler Strings
Data Logger
Data Logger
OFF
Data Log Interval
12s
Data Log Destination Log File and Printer
Select Data Log Items
Data Log Format
Text Format
Table 4-1 (Continued)
6200 Factory Default Settings