Table 4-1, 6200 factory default settings, Program installation & control – Parr Instrument 6200 User Manual
Page 20

4
Program Installation & Control
P a r r I n s t r u m e n t C o m p a n y
18
Table 4-1
6200 Factory Default Settings
Calorimeter Operations
Operating Mode
Determination
Bomb Installed/EE
1/2400.0
Heater and Pump
OFF
Operating Controls
Method of Operation
Dynamic
Reporting Units
BTU/lb
Use Spiking Correction
OFF
“OTHER” Multiplier
4.1868
Calibrate Touchscreen
LCD Backlight Timeout(s)
1200 S
LCD Backlight Intensity
70%
Print Error Messages
ON
Language
English
Spike Controls
Use Spiking
OFF
Heat of Combustion of Spike
6318.4
Use Fixed Spike
OFF
Weight of Fixed Spike
0.0
Prompt for Spike before Weight
OFF
Program Information and Control
Date & Time Settings
Volume Level Adjust
85%
Software and Hardware Info
Settings Protect
OFF
User/Factory Settings
Feature Key
Bomb Type Select
User Function Setup
Cold Restart
User/Factory Settings
User Setup ID
62-1108
Reload Factory Default Settings
Reload User Default Settings
Save User Default Settings
Calibration Data & Controls
Calibration Run Limit
10
EE Max Std Deviation
0.0
Heat of Combustion of Standard 6318.4
Bomb Service Interval
500
Control Chart Parameters
Charted Value
HOC Standard
Process Sigma
0.1
Temp Rise High Warning
3.3
Temp Rise Low Warning
2.0
Use Bomb
1
Bomb 1 Through 4
EE Value
2400.0
Protected EE Value
OFF
Thermochemical Corrections Standardization
Fixed Fuse Correction
ON 50
Acid Correction
Fixed HNO
3
10.0
Fixed Sulfur Correction
ON 0.0
Determination
Fixed Fuse Correction
ON 50
Acid Correction
Fixed HNO
3
10.0
Fixed Sulfur Correction
OFF 0.0
Net Heat/Dry Factors
Net Heat & Dry Disable
Fixed Hydrogen
OFF 0.0
Fixed Oxygen
ON 0.0
Fixed Nitrogen
ON 0.0
Calculate Net Heat of Combustion
OFF
Fixed Moisture as Determined
OFF 0.0
Fixed Moisture as Received
OFF 0.0
Dry Calculation
OFF