Menu Operating Instructions

P a r r I n s t r u m e n t C o m p a n y

44

Automatic Sample ID Controls

Automatic Sample ID: When set to ON it will

automatically assign sample identification

numbers in accordance with instructions set in

the other two keys on this menu.

Auto Sample ID Prefix: An entry here will be

used as a prefix for all sample IDs.

Next Auto Sample ID Number: Establishes the

initial sample number for a series of tests and

then shows the next sample ID which will be

assigned.

Auto Sample ID Increment: Establishes the

increment between sample numbers.

Sample Weight – Warning Above: This key displays
and leads to a sub-menu used to set the maximum
allowable sample weight (including spike) in grams.
A warning will be given if sample weights above
this value are entered.

Spike Weight Entry Mode: This key steps through
the options for entering spike weights either manu-
ally through the touch screen, automatically through
the balance (USB) port or through a network.

Sulfur Entry Mode: This key toggles steps through
the options for entering sulfur correction value
either manually through the touch screen or through
the balance (USB) port.

Moisture Entry Mode: This key steps through the
options for entering the moisture percentage wheth-
er manually through the touch screen or automati-
cally through the balance (USB) port.

Auto Preweigh ID Controls: Accesses sub-menu,
used to automatically assign sample identification
numbers when a series of samples are pre weighed
ahead of the time they are actually tested.

Preweigh Sample ID Controls

Automatic Preweigh ID: ON/OFF toggle for this

feature.

Automatic Preweigh ID Prefix: An entry here

will be used as a prefix for all pre-weigh sample

identification numbers.

Next Automatic Preweigh ID Number: Shows

the next sample identification number which will

be assigned and is used to enter the beginning

Sample ID of any series.

Automatic Preweigh ID Increment: Establishes

the increment between samples.