Hanna Instruments HI 83099 User Manual
Page 59

117
116
• Press the
Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of
HI 93750B-0 reagent. Replace the cap and gently mix for
one minute by slowly turning the cuvette upside down.
• Reinsert the cuvette into the instrument.
• Press
Timer and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and press
Read.
When the timer ends the meter will perform the reading. The
instrument displays the results in
mg/L (ppm) of potassium (K).
• Press ▲ or ▼ to access the second level functions.
• Press the
Chem Frm key to convert the result in mg/L of potassium oxide (K
2
O).
• Press ▲ or ▼ to return to the measurement screen.
INTERFERENCES
Interferences may be caused by:
Ammonium above 10 ppm
Calcium above 10000 ppm as CaCO
3
Chloride above 12000 ppm
Magnesium above 8000 ppm as CaCO
3
Sodium above 8000 ppm
Potassium MR
Potassium MR
POTASSIUM MEDIUM RANGE
SPECIFICATIONS
Range
10 to 100 mg/L
Resolution
2.5 mg/L
Accuracy
±15 mg/L ±7% of reading at 25 °C
Typical EMC
±2.5 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 610 nm
Method
Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93750
A-0
Potassium Reagent
6 drops
HI 93750
B-0
Powder Reagent
1 packet
REAGENT SETS
HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the CONCENTRATED SAMPLES procedure at page 18.
• Select the
Potassium MR method using the procedure
described in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of sample, up to the mark.
• Add six drops of HI 93750A-0, replace the cap and swirl
the solution.
• Place the cuvette into the holder and close the lid.
10 mL