63
62
• Add 12 drops of the HI 93704-0 reagent to
each cuvette. Replace the caps and shake
gently to mix (about 30 seconds).
• Place the blank (#1) into the holder and close the lid.
• Press
Timer and the display will show the countdown prior to zeroing the blank. The display will show
“
-0.0-” when the meter is zeroed and ready for measurement.
• Remove the blank.
• Insert the cuvette with the reacted sample (# 2) into the
instrument and close the lid.
• Press
Read to start the reading. The instrument displays concentration in
µµµµµ
g/L of hydrazine.
Hydrazine
# 1
# 2
# 1
# 2
INTERFERENCES
Interference may be caused by:
Highly colored samples
Highly turbid samples
Aromatic amines
Hydrazine
HYDRAZINE
SPECIFICATIONS
Range
0 to 400
µ
g/L
Resolution
1
µ
g/L
Accuracy
±4% of full scale reading at 25 °C
Typical EMC
±2
µ
g/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 420 nm
Method
Adaptation of the
ASTM Manual of Water and Environmental Technology, method
D1385-88, p-Dimethylaminobenzaldehyde method. The reaction between hydrazine
and the liquid reagent causes a yellow tint in the sample.
REQUIRED REAGENT
Code
Description
Quantity
HI 93704-0
Liquid Reagent
24 drops
REAGENT SETS
HI 93704-01 Reagents for 100 tests
HI 93704-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
• Select the
Hydrazine method using the procedure described
in the
Method Selection section (see page 12).
• Fill one cuvette up to the mark with 10 mL of distilled water.
• Place the cap, insert the cuvette # 1 into the holder and
close the lid.
• Press the
Blank key to start adjusting the light level. The display will show “Blank done” when the
meter is ready to take a zero measurement.
• Fill a second cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
10 mL
# 1
10 mL
# 2
# 1