79

78

• Fill one graduated mixing cylinder up to the 25 mL

mark with the sample.

• Add the content of one packet of HI 93730A-0

molybdenum reagent, close the cylinder and invert it
several times until completely dissolved.

• Add the content of one packet of HI 93730B-0

molybdenum reagent to the cylinder, close and invert
it several times until completely dissolved.

• Add the content of one packet of HI 93730C-0

molybdenum reagent to the cylinder, close and shake
it vigorously.

• Fill an empty cuvette with 10 mL of sample up to the

mark and replace the cap.

• Insert the cuvette into the instrument.

Molybdenum

25 mL

10 mL

Molybdenum

MOLYBDENUM

SPECIFICATIONS

Range

0.0 to 40.0 mg/L

Resolution

0.1 mg/L

Accuracy

±0.3 mg/L ±5% of reading at 25 °C

Typical EMC

±0.1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93730

A-0

Reagent A

1 packet

HI 93730

B-0

Reagent B

1 packet

HI 93730

C-0

Reagent C

1 packet

REAGENT SETS

HI 93730-01 Reagents for 100 tests
HI 93730-03 Reagents for 300 tests
For other accessories see page 132.

MEASUREMENT PROCEDURE

• Select the

Molybdenum method using the procedure

described in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to

the mark) and replace the cap.

• Place cuvette into the holder and close the lid.

• Press the

Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

10 mL