• Fill one graduated mixing cylinder up to the 25 mL
mark with the sample.
• Add the content of one packet of HI 93730A-0
molybdenum reagent, close the cylinder and invert it
several times until completely dissolved.
• Add the content of one packet of HI 93730B-0
molybdenum reagent to the cylinder, close and invert
it several times until completely dissolved.
• Add the content of one packet of HI 93730C-0
molybdenum reagent to the cylinder, close and shake
it vigorously.
• Fill an empty cuvette with 10 mL of sample up to the
mark and replace the cap.
• Insert the cuvette into the instrument.
Molybdenum
25 mL
10 mL
Molybdenum
MOLYBDENUM
SPECIFICATIONS
Range
0.0 to 40.0 mg/L
Resolution
0.1 mg/L
Accuracy
±0.3 mg/L ±5% of reading at 25 °C
Typical EMC
±0.1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 420 nm
Method
Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.
REQUIRED REAGENT
Code
Description
Quantity
HI 93730
A-0
Reagent A
1 packet
HI 93730
B-0
Reagent B
1 packet
HI 93730
C-0
Reagent C
1 packet
REAGENT SETS
HI 93730-01 Reagents for 100 tests
HI 93730-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
• Select the
Molybdenum method using the procedure
described in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
• Place cuvette into the holder and close the lid.
• Press the
Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL