Hanna Instruments HI 83099 User Manual

Page 35

background image

69

68

Iron LR

• Press the

Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cuvette.

• Fill another graduated mixing cylinder up to the 25 mL

mark with the sample.

• Add the content of one packet of HI 93746-0 TPTZ reagent,

close the cylinder and shake vigorously for 30 seconds. This
is the reacted sample.

• Fill a cuvette with 10 mL of the reacted sample up to the

mark and replace the cap.

• Insert the sample into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 30 seconds and press

Read. When the timer ends the meter will perform the reading. The

instrument displays concentration in

µµµµµ

g/L of iron.

25 mL

10 mL

Iron LR

IRON LOW RANGE

SPECIFICATIONS

Range

0 to 400

µ

g/L

Resolution

1

µ

g/L

Accuracy

±10

µ

g/L ±8% of reading at 25 °C

Typical EMC

±1

µ

g/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the TPTZ Method. The reaction between iron and the reagent causes a
violet tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93746-0

TPTZ Reagent

2 packets

REAGENT SETS

HI 93746-01 Reagents for 50 tests
HI 93746-03 Reagents for 150 tests
For other accessories see page 132.

MEASUREMENT PROCEDURE

• Select the

Iron LR method using the procedure described in

the

Method Selection section (see page 12).

• Fill one graduated mixing cylinder up to the 25 mL mark

with deionized water.

• Add the content of one packet of HI 93746-0 TPTZ reagent,

close the cylinder and shake vigorously for 30 seconds. This is
the blank.

• Fill a cuvette with 10 mL of the blank up to the mark and

replace the cap.

• Place the cuvette into the holder and close the lid.

10 mL

25 mL