93

92

Dissolved Oxygen

• Let the sample stand and the flocculent agent will start to settle.

• After approximately 2 minutes, when the upper half of the bottle becomes limpid, add

10 drops of HI 93732C-0.

• Replace the cap and invert the bottle until the settled flocculent dissolves completely.

The sample is ready for measurement when it is yellow and completely limpid.

• Fill the cuvette up to the mark with 10 mL of the unreacted

(original) sample, and replace the cap. This is the blank.

• Place the cuvette into the holder and close the lid.

• Press the

Zero key. The display will show “-0.0-” when

the meter is zeroed and ready for measurement.

• Remove the cuvette.

• Fill another cuvette up to the mark with 10 mL of the reacted

sample and replace the cap.

• Reinsert the cuvette into the instrument.

• Press

Read to start the reading. The instrument will display the results in mg/L of dissolved oxygen.

10 mL

INTERFERENCES

Interferences may be caused by reducing and oxidizing materials.

Dissolved Oxygen

SPECIFICATIONS

Range

0.0 to 10.0 mg/L

Resolution

0.1 mg/L

Accuracy

±0.4 mg/L ±3% of reading at 25 °C

Typical EMC

± 0.1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, Azide modified Winkler method. The reaction between dissolved oxygen

and the reagents causes a yellow tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93732

A-0

Reagent A

5 drops

HI 93732

B-0

Reagent B

5 drops

HI 93732

C-0

Reagent C

10 drops

REAGENT SET

HI 93732-01 Reagents for 100 tests
HI 93732-03 Reagents for 300 tests
For other accessories see page 132.

MEASUREMENT PROCEDURE

• Select the

Dissolved Oxygen method using the procedure described in

the

Method Selection section (see page 12).

• Fill one 60 mL glass bottle completely with the unreacted sample.

• Replace the cap and ensure that a small part of the sample spills

over.

• Remove the cap and add 5 drops of HI 93732A-0 and 5 drops of

HI 93732B-0.

• Add more sample, to fill the bottle completely. Replace the cap

again and ensure that a part of the sample spills over. This is to
make sure that no air bubbles have been trapped inside, which
could alter the reading.

• Invert several times the bottle. The sample becomes orange-yellow

and a flocculent agent will appear.

DISSOLVED OXYGEN