Phosphate HR
• Add the content of one packet of HI 93717B-0 Phosphate
HR Reagent B to the cuvette. Replace the cap and shake
gently until completely dissolved.
• Reinsert the cuvette into the instrument.
• Press
Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 5 minutes and press
Read. When the timer ends the meter will perform the reading. The instrument
displays the results in
mg/L of phosphate (PO
4
3
¯).
• Press ▲ or ▼ to access the second level functions.
• Press the
Chem Frm key to convert the result in mg/L of phosphorus (P) and phosphorus pentoxide
(P
2
O
5
).
• Press ▲ or ▼ to return to the measurement screen.
INTERFERENCES
Sulfide
Chloride above 150000 mg/L
Calcium above 10000 mg/L as CaCO
3
Magnesium above 40000 mg/L as CaCO
3
Ferrous iron above 100 mg/L
Phosphate HR
PHOSPHATE HIGH RANGE
SPECIFICATIONS
Range
0.0 to 30.0 mg/L
Resolution
0.1 mg/L
Accuracy
±1 mg/L ±4% of reading at 25 °C
Typical EMC Dev. ±0.1 mg/L
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition, Amino Acid method. The reaction between phosphate and reagents
causes a blue tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93717
A-0
Molybdate
10 drops
HI 93717
B-0
Reagent B
1 packet
REAGENT SETS
HI 93717-01 Reagents for 100 tests
HI 93717-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
• Select the
Phosphate HR method using the procedure described
in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the
Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add 10 drops of HI 93717A-0 Molybdate reagent.
10 mL