Iron HR

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively,

wait for 3 minutes and press

Read. When the timer ends the meter will perform the reading.

• The instrument displays concentration in

mg/L of iron.

INTERFERENCES

Interference may be caused by:
Molybdate Molybdenum above 50 ppm
Calcium above 10000 ppm (as CaCO

3

)

Magnesium above 100000 ppm (as CaCO

3

)

Chloride above 185000 ppm.

Iron HR

IRON HIGH RANGE

SPECIFICATIONS

Range

0.00 to 5.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.04 mg/L ±2% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the

EPA Phenantroline method 315B, for natural and treated waters.

The reaction between iron and reagents causes an orange tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93721-0

Powder Reagent

1 packet

REAGENT SETS

HI 93721-01 Reagents for 100 tests
HI 93721-03 Reagents for 300 tests
For other accessories see page 132.

MEASUREMENT PROCEDURE

• Select the

Iron HR method using the procedure described

in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to the

mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the

Zero key. The display will show “-0.0-” the meter is zeroed and ready for measurement.

• Remove the cuvette and add the content of one packet

of HI 93721-0 reagent. Replace the cap and shake
until dissolution is complete.

10 mL