Color of Water

# 3

• Fill the second cuvette up to the mark with unfiltered sample

and replace the cap. This is the apparent color.

• Filter 10 mL of sample through a filter with a 0.45

µ

m

membrane into the third cuvette, up to the 10 mL mark and
replace the cap. This is the true color.

• Insert the apparent color cuvette (# 2) into the instrument and

close the lid.

• Press

Read to start the reading.

• The meter displays the value

of apparent color in PCU.

• Remove the cuvette, insert the true color cuvette (# 3) into the

instrument and ensure that the notch on the cap is positioned
securely into the groove.

• Press

Read to start the reading. The meter displays the value of true color in PCU.

# 2

# 3

10 mL

Color of Water

COLOR OF WATER

SPECIFICATIONS

Range

0 to 500 PCU (Platinum Cobalt Units)

Resolution

1 PCU

Accuracy

±10 PCU ±5% of reading at 25 °C

Typical EMC

± 1 PCU

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, Colorimetric Platinum Cobalt method.

REQUIRED ACCESSORIES

0.45

µ

m membrane for true color measurement.

For other accessories see page 132.

MEASUREMENT PROCEDURE

• Select the

Color of Water method using the procedure described in

the

Method Selection section (see page 12).

• Fill one cuvette up to the mark with deionized water and replace

the cap. This is the blank.

• Place the blank (# 1) into the holder and close the lid.

• Press the

Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the blank.

# 1

# 1

10 mL

# 2