75

74

Manganese LR

MANGANESE LOW RANGE

SPECIFICATIONS

Range

0 to 300

µ

g/L

Resolution

1

µ

g/L

Accuracy

±10

µ

/L ±3% of reading at 25 °C

Typical EMC

±1

µ

g/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the PAN Method. The reaction between manganese and the reagents
causes an orange tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93748

A-0

Ascorbic acid

2 packets

HI 93748

B-0

Alkaline-cyanide sol.

0.40 mL

HI 93748

C-0

0.1% PAN indicator

2 mL

HI 93703-51

Dispersing Agent

4-6 drops

REAGENT SETS

HI 93748-01 Reagents for 50 tests
HI 93748-03 Reagents for 150 tests
For other accessories see page 132.

MEASUREMENT PROCEDURE

• Select the

Manganese LR method using the procedure described in the Method Selection section (see

page 12).

• Fill one cuvette with 10 mL of deionized water (up to the mark).

• Fill a second cuvette with 10 mL of sample (up to the mark).

• Add the content of one packet of HI 93748A-0 Ascorbic acid to

each cuvette, replace the caps and shake gently until completely
dissolved.

10 mL

10 mL

Manganese HR

• Add the content of one packet of HI 93709B-0 Sodium

Periodate reagent. Replace the cap and invert to mix with
gently movements for 2 minutes.

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 1 minute and 30 seconds and press

Read. When the timer ends the meter will perform the reading.

The instrument displays the results in

mg/L of manganese.

• Press ▲ or ▼ to access the second level functions.
• Press the

Chem Frm key to convert the result in mg/L of potassium permanganate (KMnO

4

) and

permanganate (MnO

4

¯).

• Press ▲ or ▼ to return to the measurement screen.

INTERFERENCES

Interference may be caused by:
Calcium above 700 mg/L
Chloride above 70000 mg/L
Iron above 5 mg/L
Magnesium above 100000 mg/L.