Hanna Instruments HI 83099 User Manual
Page 33

65
64
Iodine
• Reinsert the cuvette into the instrument.
• Press
Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 2 minutes and 30 seconds and press
Read. When the timer ends the meter will perform the
reading.
• The instrument displays concentration in
mg/L of iodine.
INTERFERENCES
Interference may be caused by: Bromine, Chlorine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately
2 minutes after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or acidity greater than 150 mg/L CaCO
3
,
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.
IODINE
SPECIFICATIONS
Range
0.0 to 12.5 mg/L
Resolution
0.1 mg/L
Accuracy
±0.1 mg/L ±5% of reading at 25 °C
Typical EMC
±0.1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition, DPD method. The reaction between iodine and the reagent causes a pink
tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93718-0
DPD Reagent
1 packet
REAGENT SETS
HI 93718-01 Reagents for 100 tests
HI 93718-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
• Select the
Iodine method using the procedure described in
the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the
Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cap and add the content of one packet of
HI 93718-0 DPD reagent. Replace the cap and shake
gently for about 30 seconds to dissolve most of the reagent.
10 mL
Iodine