Cleaning procedure

General

Soak in Hanna HI 7061 General Cleaning

Solution for approximately ½ hour.

Removal of films, dirt or deposits on the membrane/junction:
Protein

Soak in Hanna HI 7073 Protein Cleaning
Solution for 15 minutes.

Inorganic

Soak in Hanna HI 7074 Inorganic Cleaning

Solution for 15 minutes.

Oil/grease

Rinse with Hanna HI 7077 Oil and Fat
Cleaning Solution.

IMPORTANT: After performing any of the cleaning procedures rinse the

probe thoroughly with distilled water and soak the electrode in
HI 70300 Storage Solution for at least 1 hour before

reinstalling it.

Troubleshooting

Evaluate your probe performance based on the following.

• Noise (Readings fluctuate up and down) could be due to

clogged or dirty junction: refer to the Cleaning Procedure
above.

• Dry Membrane/Junction: soak in Storage Solution HI 70300

for at least 1 hour. Check to make sure the installation is
such as to create a well for the probe bulb to constantly
remain moist.

• Drifting: soak the probe tip in warm Hanna Solution HI

7082 for one hour and rinse tip with distilled water.

• Low Slope: refer to the cleaning procedure above.

• No Slope:

- Check the probe for cracks in glass stem or bulb (replace

the probe if cracks are found).

- Make sure cable and connections are not damaged.

• Slow Response/Excessive Drift: soak the tip in Hanna Solu-

tion HI 7061 for 30 minutes, rinse thoroughly in distilled
water and then follow the Cleaning Procedure above.

• For ORP probe: polish the metal tip with a lightly abrasive

paper (paying attention not to scratch the surface) and wash
thoroughly with water.