A 1/2 BSP sample input fitting allows direct connection to the
optional input filter.
Sample line pressure should be between 0.07 and 4 bar (1
and 57.2 psi) with an ideal pressure of 0.7 bar (10 psi).
It is recommended to assure that the water inlet come at 1 m
above the instrument sample input pitting. For maintenance
issues is also recommended to install valves on both sam-
pling pipes.
Drain Line Installation
The drain hose fitting is a 20 mm (3/4”) hose barb on the
bottom of the instrument enclosure. An air gap between the end
of the drain hose and the drain is recommended to prevent any
back flow into the instrument in the event of drain blockage.