STARTUP

To power up the analyzer open the electronic
box door and turn on the main switch.

When the analyzer is powered up, the display
backlight is turned on and the initialization take
place. In this phase, the integrity of the stored
data is checked and the information regarding
the language is loaded.

The display will show HANNA INSTRUMENTS, the name of
the instrument and the software version.

Note:

If the instrument is set for free chlorine analysis, the software
will report at startup Free Chlorine and if it is set for total
chlorine, the software will report at startup Total Chlorine.

After initialization, the analyzer will show the main panel (or
chlorine measuring panel for PCA 310). The measured value
are displayed. The chlorine concentration will be updated only
after a full measuring cycle. The first reading is 0.00 mg/L and
the dosing relay is not active.

After the first chlorine concentration is measured and displayed,
the chlorine dosing relay is activated if necessary.

Note:

If the SMS feature is selected and correctly configured, the
analyzer will send a SMS at each power up sequence.