GSM

GSM MODE

If the “Type” is set to GSM, the analyzer will work with HI 504900
GSM module. This connection enables the analyzer to send
SMSs to one (or two) cellular phone(s) and through this feature
the device can be monitored. Moreover if an error occurs on
the PCA 3x0, a SMS is sent to the cellular phone(s) signaling
immediately the user about the problem.

When the SMS feature is active, three type of SMS messages
are sent to the selected cellular phone(s).

Alarm SMS which contains the Error, Alarm and Warning messages

that are active on the analyzer, separated by ‘;’. The events
that generate the alarm SMS are user selectable from “Gen-
eral Menu” - “SMS Settings”.

Info SMS which contains the Chlorine, pH, ORP and Temperature read-

ings. Analyzer’s error status and settings are also included as
selected by the user from “General Menu” - “SMS Settings”.

Warning SMS which contains information about the SIM card credit.

The settings related to GSM are grouped in “General Menu”
- “Serial & GSM Comm.”.

SETTING THE GSM FEATURE

To use the SMS feature, a SIM card able to make voice calls
must be used.

It is necessary to enter
the PIN code of the
SIM card inserted in
the GSM module. To
do this, set the “Gen-
eral Menu” - “Serial
& GSM Comm” -
“PIN No.”.

Note:

The PIN number is not
visible after editing for
protection reasons.
The field will always
show 0000.

Type :STANDARD
Baud Rate:9600
Address: 01
Send SMS :Inactive

PIN No.

:0000

Phone 1

:Inactive

No1:---------------
Phone 2

:Inactive

No2:---------------
Charge Inf:Inactive
Exp.Date:2010/01/01
Remaining:0000
RepeatNo:2
Delay

:05

min

Remote

:Active

This manual is related to the following products: