Principle of operation – Hanna Instruments HI 84529 User Manual

Page 8

background image

www.

hannainst

.com

8

Description

In an ideal solution, the endpoint of an acid titration corresponds

stoichiometrically to the complete neutralization of the acids present.

This endpoint for this reaction can be determined visually using a color

indicator (i.e. phenolphthalein), however this endpoint becomes very

objective in opaque and colored samples. The HI 84529 removes this issue

by titrating to a fixed pH endpoint that can be set by the user. The HI 84529

can be customized to meet individual needs with samples titrated by mass or

by volume, in two different ranges, with diluted and non diluted (low range

only). For precise analysis the sample size, volume of titrant added and titrant

concentration must be known.

The HI 84529 Titratable Acid in Dairy Products Minititrator utilizes a simple

sample preparation, a high quality dosing pump for titrant additions,

potentiometric endpoint determination and instantaneous computations.

To maintain the high accuracy of the minititrator a simple pump calibration

is required. The pump calibration uses a known quantity of a known solution

to compensate for changes in the dosing system, this procedure should be

performed regularly.

Principle of

Operation