Operational Guide

This screen appears when the stirrer is

not working properly. Check the stir bar

and beaker content. Press RESTART to

try again.

This error message appears when the

pump is not working properly. Check the

tubing, valve and syringe. Press Restart

to try again.

The instructions listed below should be followed carefully to ensure

measurements are conducted with the highest possible accuracy and precision.

IT IS CRITICAL THAT THE TIP BE IMMERSED IN THE SOLUTION BEING TITRATED

(APPROXIMATELY 0.25 CM).

• Calibrate the pump prior to each series

of titrations.

• Calibrate the pump if the meter is left

idle for several hours.

• Analyze the sample immediately after

it is obtained.

• Clean the electrode with HI 70640

cleaning solutions specially designed

for milk industry.

View/Delete Titrator Recorded Data

Press MENU key then Recall to access the

titrator logs.

When an external USB storage device is

connected, the Export key is displayed. It

saves the meter and titrator logs in two

Titration

Procedure

Tips for an

Accurate

Measurement