Maintenance

Symptoms

Problem

Solution

Slow response/excessive

drift.

Dirty pH or reference

electrode.

Soak the electrode tip HI 70640

solution for 30 minutes and

follow the cleaning procedure.

Refill reference electrode with

new fill solution.

Reading fluctuates up and

down (noise).

Clogged/dirty junction.

Low electrolyte level

(reference electrode).

Cable connection.

Clean the pH electrode. Refill

reference electrode with fresh fill

solution. Check cable connection

to meter and verify protective

cap is off.

While in pH reading mode,

-2.00 or 16.00 pH is displayed

blinking.

Reading out of range.

Check cable connection to meter

and verify protective cap is off.

Check the quality of the sample.

Clean the electrodes. Refill with

fresh fill solution.

The meter does not accept

the pH buffer solution for

calibration.

Broken pH or reference

electrode.

Follow the electrode cleaning

procedure. If the error persists

replace the electrode or contact

your local Hanna Representative.

The pump calibration can't be

performed.

Broken pump tubing.

Wrong or contaminated

pump calibration solution.

Broken pH electrodes.

Verify tubing, valve, syringe

are intact and solution passes

when pump is primed and no air

bubbles are present. Check the

pump calibration solution. Verify

electrodes are calibrated. Prepare

another standard, prime the

pump and restart the calibration.

The temperature probe is

connected, but the meter

displays "MTC".

Broken temperature

probe.

Replace temperature probe.

Troubleshooting

Guide