Operational Guide

Note: The aspiration tube must be inserted in the titrant bottle. The dosing

tip must be placed over a rinse beaker.

Before starting the prime procedure, make sure you are using the appropriate

titrant solution for the selected range.

To prime the burette, select PRIME option

from TITRATION mode.

Adjust the rinses number by pressing the

‚

and  keys and press START.

The number of syringe rinses can be set

between 1 and 5 (at least three rinses

are recommended to ensure that the air

bubbles are completely removed).

To pause the prime process press

the PAUSE key; to continue press the

CONTINUE key. To stop the prime process

press the STOP key.

Dosing

Pump Prime

Procedure