Electrode calibration procedure – Hanna Instruments HI 84529 User Manual

Page 27

background image

www.

hannainst

.com

27

Operational Guide

and press STIR. If necessary press the ARROW keys in order to select a

different buffer value.

4. The (unstable measurement) symbol will be shown on the display until

the reading becomes stable.

5. When the reading is stable and close to the selected buffer, the

(unstable measurement) symbol will disappear and the CONFIRM key

will become active.

6. Press CONFIRM to confirm the calibration. The instrument stores the

calibration value and returns to calibration menu, where the date and

time for the pH calibration will be updated.

Note: A buffer confirmed during the calibration process is removed from

the list of available buffers. If the value measured by the instrument is

not close to the selected buffer, a “Wrong Buffer” error message will

be shown on the display. Check if the correct buffer has been used or

regenerate the pH electrodes by following the Cleaning Procedure. If

necessary change the buffer or the electrode.

If the measured offset isn’t within the preset limits (±45 mV), the meter will

display the message “Buffer Contaminated” alternatively with ”Electrode

Dirty/Broken”.

If the computed slope isn’t within the

preset limits, the meter will display the

message “Wrong Slope”. If the slope is too
high the

symbol will be displayed. If

the slope is too low the

symbol will

be displayed.

Electrode

Calibration

Procedure